Ultrasound physics

What the machine is doing while you look for the nerve. Watch the film, then try the controls yourself.

1 min 36 s · animated, with on-screen captions · part 1 of the Basics series

The essentials

Six ideas that explain the picture

01

Pulse and echo

The probe sends a short pulse and then listens. The machine assumes a speed of sound of 1540 m/s in soft tissue and places each echo by the time it took to return: depth = speed × time ÷ 2.1

02

Frequency

Higher frequency gives a shorter wavelength and better resolution, but is attenuated sooner. Use the highest frequency that still reaches the target.1

03

Echogenicity

Echoes form where tissues differ in acoustic impedance. Strong reflectors appear bright (hyperechoic), weak ones dark (hypoechoic), and fluid black (anechoic).1

04

Anisotropy

Nerves and tendons reflect best when the beam meets them at 90°. A small tilt of the probe can make a nerve fade or reappear.2

05

Attenuation

Sound loses energy with depth, more so at higher frequencies. Time-gain compensation amplifies the later echoes to even out the image.1

06

Artefacts

Acoustic shadowing behind bone, enhancement behind fluid and reverberation beneath a needle all arise when tissue breaks the machine's assumptions.2

Try it

Tilt the probe and watch the nerve

Move the tilt slider a few degrees. The fascial lines and the nerve fade while the artery stays black.

Schematic teaching image drawn by the site. It is not a patient scan.

Next: Knobology

References

  1. Sites BD, Brull R, Chan VWS, et al. Artifacts and pitfall errors associated with ultrasound-guided regional anesthesia. Part I: understanding the basic principles of ultrasound physics and machine operations. Reg Anesth Pain Med 2007; 32: 412-8. Review
    https://doi.org/10.1016/j.rapm.2007.05.005
  2. Sites BD, Brull R, Chan VWS, et al. Artifacts and pitfall errors associated with ultrasound-guided regional anesthesia. Part II: a pictorial approach to understanding and avoidance. Reg Anesth Pain Med 2007; 32: 419-33. Review
    https://doi.org/10.1016/j.rapm.2007.08.001